CALL[:CELL]:RFGenerator

CALL
[:CELL]
"Diagram Conventions" on page 168
CALL[:CELL]:RFGenerator:FREQuency
Function
Sets/queries the RF generator downlink channel frequency. If no units are specified the setting
defaults to Hz.
The test set will tune to the frequency entered using this command when the
"CALL[:CELL]:CONTrol:DOWNlink:FREQuency:AUTO"
Setting
Range: 292 MHz to 2700 MHz
Resolution: 0.1 Hz
Query
Range: 292 MHz to 2700 MHz
Resolution: 0.1 Hz
*RST Setting
851 MHz
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:CELL:RFGENERATOR:FREQUENCY 450 MHZ"
!Sets the downlink channel frequency to 450 MHz.
:RFGenerator
:FREQuency
<sp><num value>[HZ|KHZ|MHZ|GHZ]
? (returns num value)
command is set to manual.