Operation

Setup and Measurement Procedure

Setup and measurement procedure is as follows:
1. Connect the 16048E directly to the UNKNOWN terminals and GUARD
terminal of the instrument.
2. Connect the user-fabricated test fixture to the test leads.
3. Perform Open/Short compensation as described in the instrument's operation
manual.
4. Connect the DUT to the 16048E test fixture.
Connect the green lead to the chassis or grounded part on the test fixture to reduce
NOTE
the effects of stray capacitance between the DUT and the test fixture.

BNC Bracket

A BNC bracket (Agilent PN 16032-60001) is furnished with the 16048E. Figure
1-2 shows an example for using the BNC bracket with the 16048E. By connecting
the 16048E to the BNC bracket, the 16048E's outer shield conductors are
connected together to construct the four-terminal pair measurement circuit
configuration. For further details about the four-terminal pair measurement refer to
the instrument's Operation Manual.
Figure 1-2
BNC bracket application
Chapter 1