SETup:AFANalyzer
SETup:AFANalyzer:COUNt:NUMBer
Function
Sets/queries the number of AF analyzer measurements to be made when multi-measurement
state is ON.
Setting
Range: 1 to 999
Query
Range: 1 to 999
*RST Setting
10
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:COUNT:NUMBER 5"
!Sets the number of measurements to 5 when the multi-measurement state is set to ON.
SETup:AFANalyzer:COUNt:STATe
Function
Sets/queries the multi-measurement state of the AF analyzer measurement.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:COUNT:STATE ON" !Turns on multi-measurements.
SETup:AFANalyzer:DEMPhasis:STATe
Function
Sets/queries whether 750 us de-emphasis is enabled.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:DEMPHASIS:STATE ON"