Table 2. Partial Preset Behavior
Function
Maskable Message Display
State
Operating Mode
Measurements
Measurement results
Measurement integrity
indicator
Transmit Power State
Cell Activated State
Call in progress
Call Control Status
Call Counters
Call Error Counters
SACCH
Pending Service request
Full Preset
A full preset requires you to select new measurements and configure their parameters. If measurement
parameters do not need to be changed, use a partial preset to save time.
Example
OUTPUT 714;"*RST" !Recommended command for a full preset, sets trigger arm to single.
OUTPUT 714;"SYSTEM:PRESET2" !Command for a full preset, sets trigger arm to continuous.
The *RST common command is the recommended command for a full preset operation.
NOTE
Transmit power is not set to OFF during a full preset, transmit power is set to the default value.
Partial Preset Result
no change
Active Cell
aborted and inactivated
NAN
1 = no result available
ON
ON
aborted
Idle
cleared
cleared
cleared
not cleared
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_gen_inst_states.fm
Preset Descriptions
117