SYSTem:CORRection:SFRequency
Function
Sets/queries up to 20 comma separated frequency values in MHz. This command sets the state to
On.
The units (GHz, MHz, kHz, Hz) are optional. If no units are specified, units default to MHz.
See
"Amplitude Offset" on page
Sending a null list (no values) sets the state of all frequencies to Off.
Setting
Range: 0 to 20 comma separated values from 292.5 MHz to 2700 MHz
Resolution: 1 Hz
Query
Range: 1 to 20 comma separated values from 292.5 MHz to 2700 MHz, and NAN (9.91E+37)
Resolution: 1 HZ
Factory
1. 800.00 MHz Off
Setting
2. 810.00 MHz Off
3. 820.00 MHz Off
4. 830.00 MHz Off
5. 840.00 MHz Off
6. 850.00 MHz Off
7. 860.00 MHz Off
8. 870.00 MHz Off
9. 880.00 MHz Off
10. 890.00 MHz Off
11. 900.00 MHz Off
12. 910.00 MHz Off
13. 920.00 MHz Off
14. 930.00 MHz Off
15. 940.00 MHz Off
16. 950.00 MHz Off
17. 960.00 MHz Off
18. 970.00 MHz Off
19. 980.00 MHz Off
20. 990.00 MHz Off
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SYSTEM:CORRECTION:SFREQUENCY 1710.2 MHZ,1805.2 MHZ,1784.8 MHZ,1879.8 MHZ"
!Sets up to 20 comma separated frequencies and sets the state to On for the frequency,
!offset, and RF IN/OUT Amplitude Offset State
OUTPUT 714;"SYSTEM:CORRECTION:SFREQUENCY"
!Sets the state for all frequencies and offsets to Off