SETup:AFANalyzer:DETector[:TYPE]
Function
Sets/queries the detector type for audio level measurements.
Setting
Range: RMS|PEAK
Query
Range: RMS|PEAK
*RST setting
RMS
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:DETECTOR:TYPE PEAK"
SETup:AFANalyzer:EXPandor[:SRLevel]
Function
Sets/queries the expandor's reference level and automatically sets the expandor's state to ON.
Setting
Range: 10 mV to 10 V
Resolution: 1 mV
Query
Range: 10 mV to 10 V
*RST setting
1 V
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:EXPANDOR:SRLEVEL 5V"
!Sets expandor state to on and level to 5 V.
SETup:AFANalyzer:EXPandor:RLEVel
Function
Sets/queries the expandor's reference level to be used when the expandor's state is set to ON.
Setting
Range: 10 mV to 10 V
Resolution: 1 mV
Query
Range: 10 mV to 10 V
*RST setting
1 V
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:EXPANDOR:RLEVEL 5V" !Sets expandor reference level to 5 V.