Agilent Technologies 8960 Series Reference Manual page 257

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

FETCh:WWQuality:OOFFset:ALL?
Function
Queries the origin offset measurements.
Results are returned in the following order: minimum, maximum, average, standard deviation
origin offset .
Query
Minimum Origin Offset
• Range: -99 to 0 dB, 9.91 E+37
• Resolution: 0.1
Maximum Origin Offset
• Range: -99 to 0 dB, 9.91 E+37
• Resolution: 0.1
Average Origin Offset
• Range: -99 to 0 dB, 9.91 E+37
• Resolution: 0.1
Standard Deviation Origin Offset
• Range: 0 to 99 dB, 9.91 E+37
• Resolution: 0.01
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WWQUALITY:OOFFSET:ALL?"
!Returns mininum, maximum, average, and standard deviation origin offset.
FETCh:WWQuality:OOFFset:AVERage?
Function
Queries the average origin offset .
Query
Range: -99 to 0 dB, 9.91 E+37
Resolution: 0.1
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WWQUALITY:OOFFSET:AVERAGE?" !Returns average origin offset.
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_fetch_wwquality.fm
FETCh:WWQuality