Agilent Technologies 8960 Series Reference Manual page 291

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

SETup:AFANalyzer:FREQuency:STATe
Function
Sets/queries whether the audio frequency measurement is enabled.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:FREQUENCY:STATE ON"
SETup:AFANalyzer:PEAK:VOLTage
Function
Sets/queries the expected peak voltage for audio measurements.
Setting
Range: 0.001 to 20 Vpk
Resolution: 0.001
Query
Range: 0.001 to 20 Vpk
*RST setting
20
Requirements
Test Application Revision: A.02.00 and above
Programming Example
10 OUTPUT 714;SETUP:AFANALYZER:PEAK:VOLTAGE 10V"
SETup:AFANalyzer:SDIStortion:FREQuency
Function
Sets/queries the fundamental frequency setting for the SINAD/distortion measurement.
Setting
Range: 100.0 to 10000 Hz
Resolution: 0.1
Query
Range: 100.0 to 10000 Hz
*RST setting
1000
Requirements
Test Application Revision: A.02.00 and above
Programming Example
10 OUTPUT 714;SETUP:AFANALYZER:SDISTORTION:FREQUENCY 300HZ"
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_setup_afan.fm
SETup:AFANalyzer
291