SETup:AFANalyzer
SETup:AFANalyzer:SDIStortion:STATe
Function
Sets/queries whether the SINAD/distortion measurement is enabled.
Setting
Range: 1|0N|0|OFF
Query
Range: 1|0
*RST setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
10 OUTPUT 714;SETUP:AFANALYZER:SDISTORTION:STATE ON"
SETup:AFANalyzer:TIMeout[:STIMe]
Function
Sets/queries the AF analyzer measurement's timeout setting and automatically sets the timeout
state to ON.
Setting
Range: 1 to 999 seconds
Resolution: 1 second
Query
Range: 1 to 999
*RST Setting
10 seconds, on
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:TIMEOUT:STIME 5S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
SETup:AFANalyzer:TIMeout:STATe
Function
Sets/queries the AF analyzer measurement's timeout state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:AFANALYZER:TIMEOUT:STATE ON"
292