CALL:STATus
CALL:STATus:CCPChannel:PRIMary[:LEVel]?
Function
Queries the current level of the primary common control physical channel.
Query
Range: -20.00 to 0 dB, 9.91 E+37
Resolution: 0.01 dB
*RST Setting
9.91 E+37
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:CCPCHANNEL:PRIMARY:LEVEL?"
CALL:STATus:CCPChannel:PRIMary:STATe?
Function
Queries the current state of the primary common control physical channel.
Query
Range: 0|1
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:CCPCHANNEL:PRIMARY:STATE?"
CALL:STATus:CELL:POWer[:AMPLitude]?
Function
Queries the current cell power level.
Cell power is set using the
Query
Range:
• -170dBm/3.84 MHz to +37dBm/3.84 MHz, 9.91 E+37
(-120 dBm/3.84 MHz to -13 dBm/3.84 MHz actual hardware range with amplitude offset = 0)
The actual power range is defined by adding the value of the associated amplitude offset to
the range above. For example, if amplitude offset is -3.5 dB, then the power range will be
-123.5 dBm/3.84 MHz to -16.5 dBm/3.84 MHz.
Resolution: 0.01 dBm/3.84 MHz
*RST Setting
9.91 E+37
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:CELL:POWER:AMPLITUDE?"
210
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_call_status.fm
"CALL[:CELL]:POWer"
commands.