INITiate
INITiate:DONE:FLAG[:ALL]
Function
Specifies that all measurements are either included or excluded when the INITitate:DONE?
query is sent.
If a measurement trigger arm is left in continuous mode, the done flag for that measurement will
toggle between DONE and WAIT. The INITiate:DONE? query will probably not be able to catch
the measurement at the instant it is done; therefore, the measurement will never appear to be
done. If a measurement's triggering must be continuously armed, you should exclude the
measurement using
Use INITitate:DONE:FLAG:ALL EXCLude to exclude all measurements.
Once INITiate:DONE:FLAG has been set to EXCLude for a measurement, the user must send
the INCLude command for that measurement in order to query that measurement with the
INITiate:DONE? query. The test set will not reset any excluded measurement with any type of
preset.
Setting
Range: INCLude|EXCLude
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"INITIATE:DONE:FLAG:ALL INCLUDE"
!Includes all measurements in the INITIATE:DONE? query.
INITiate:DONE:FLAG:<measurement mnemonic>
Function
Specifies if a measurement done flag is considered (using the INCLude or EXCLude settings)
when the INITitate:DONE? query is sent.
If more than one measurement is active, and one or more measurement trigger arm states is
continuous, the INITiate:DONE? query may not be effective. The best solution is to change the
arm state of all measurements to single using the
However, if a measurement must be continuously triggered, you should exclude the measurement
using the EXCLude setting provided by this command if you want to use INITiate:DONE? to
query measurement done status.
The test set will not reset any excluded measurement to included with any type of preset.
Setting
Range: INCLude|EXCLude
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"INITIATE:DONE:FLAG:WAP EXCLUDE"
!Excludes the average power measurement from responding to the INITIATE:DONE? query.
266
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_initiate.fm
"INITiate:DONE:FLAG:<measurement mnemonic>"
.
"SETup[:ALL]:CONTinuous"