Step 3: Set Measurement Parameters
NOTE
You can set the Trigger Arm for all measurements to single (CONTinuous OFF) using the
following command syntax:
Example 3. Programming Example:
OUTPUT Test_set;"SET:CONT:OFF"
sets the trigger arm to single for all measurements.
Example 4. Programming Example:
OUTPUT Test_set;"SET:DTXP:TRIG:SOUR AUTO"
sets the trigger source to AUTO for the digital transmit power measurement.
Example 5. Programming Example:
OUTPUT Test_set;"SET:DTXP:TRIG:DEL 10 US"
sets the trigger delay to 10 microseconds for the digital transmit power measurement.
Measurement-Specific Measurement Parameters
Measurement-specific measurement parameters set up operating conditions for a specific measurement. For
example:
• Setting the channel power measurement speed in IS-2000
• Setting ORFS frequency offsets in GPRS
• Setting number of bits for the FBER measurement in GSM
Refer to the GPIB syntax listing for the detailed list of measurement parameters for individual
measurements.
Example 6. Programming Example:
OUTPUT Test_set;"SET:CPOW:MSP FAST"
sets the IS-2000 channel power measurement speed to fast.
Example 7. Programming Example:
OUTPUT Test_set;"SET:ORFS:SWIT:FREQ 400 KHZ"
sets the first ORFS offset to 400 kHz.
Example 8. Programming Example:
OUTPUT Test_set;"SET:FBER:COUN 10000"
sets the number of fast BER bits to test to 10,000 bits.
82
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_prog_generic_step3.fm
SETup[:ALL]:CONTinuous:OFF