CALL:MS:POWer

:MS
:POWer
CALL
"Diagram Conventions" on page 168
CALL:MS:POWer:TARGet[:AMPLitude]?
Function
Sets/queries the UE (mobile station) target transmitter power.
When operating in FDD test mode, this setting sets the UE expected power into the test set when
using automatic receiver control. This setting does not control the UE output power in any way.
The user must set this value when making measurements that do not provide autoranging to
prevent an over range or an under range measurement situation. See
Mode"
When Active Cell operating mode is made available (future test application release): If the uplink
closed loop power control is in the active bits mode, the instrument will measure the mobile's
transmit power and it will send the required transmit power control bit patterns to the mobile.
This setting does not directly set the transmit power of the mobile, but it will drive the mobile
power to the setting if closed loop power is active. As in FDD Test operating mode, this setting
has the secondary function of setting the test set's receiver power level when the receiver power
control state is set to AUTO.
Query
Range: -50 dBm to +24 dBm
*RST Setting
24 dBm
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:MS:POWER:TARGET:AMPLITUDE -20"
:TARGet
[:AMPLitude]
.
<sp><num value>[DBM]
?(returns num value)
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_call_ms_power.fm
CALL:MS:POWer
"FDD Test Operating