Measurement Calibration

This measurement should be calibrated using the Calibrate Measurements function (CAL:MEAS?) under any
of the following conditions:
• The E1963A W-CDMA test application was just activated and this measurement has not been calibrated
yet. If you were using a different test application previous to activating the W-CDMA test application, or if
you have just received your Agilent 8960 Series 10 test set with the W-CDMA test application installed, you
must calibrate this measurement before using it.
• The temperature has changed by ± 10° C or more since the last calibration. If this situation exists, the
integrity indicator value is changed and a message is displayed indicating "Uncalibrated Due to
Temperature".
• One year has elapsed since the last calibration.

Related Topics

"Programming an Adjacent Channel Leakage Ratio Measurement"
"Adjacent Channel Leakage Ratio Troubleshooting"
"Test Adherence to Standards"
"Triggering of Measurements"
"Statistical Measurement Results"
"Integrity Indicator"
Adjacent Channel Leakage Ratio Measurement Description