Measurement Calibration
This measurement should be calibrated using the Calibrate Measurements function (CAL:MEAS?) under any
of the following conditions:
• The E1963A W-CDMA test application was just activated and this measurement has not been calibrated
yet. If you were using a different test application previous to activating the W-CDMA test application, or if
you have just received your Agilent 8960 Series 10 test set with the W-CDMA test application installed, you
must calibrate this measurement before using it.
• The temperature has changed by ± 10° C or more since the last calibration. If this situation exists, the
integrity indicator value is changed and a message is displayed indicating "Uncalibrated Due to
Temperature".
• One year has elapsed since the last calibration.
Related Topics
•
"Programming an Adjacent Channel Leakage Ratio Measurement"
•
"Adjacent Channel Leakage Ratio Troubleshooting"
•
"Test Adherence to Standards"
•
"Triggering of Measurements"
•
"Statistical Measurement Results"
•
"Integrity Indicator"
Adjacent Channel Leakage Ratio Measurement Description