SYSTem:APPLication
SYSTem:APPLication:CATalog:REVision:COUNt? '<test application name>'
Function
Query the test set for the number of revisions present on the hard disk for a specified test
application. Up to 30 revisions can be stored for a test application.
The test application name must be entered as it appears in the test application Setup menu, with
the exception that the string is not case sensitive and can be entered in any combination of upper
and lower case letters.
Query
Range: 0 through 30
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SYSTEM:APPLICATION:CATALOG:REVISION:COUNT? 'AMPS/136 MOBILE TEST'"
SYSTem:APPLication:FORMat[:NAME]
Function
Switches formats in a fast switching test application. A licensed fast switching test application
must be selected to use this command.
Queries the test set for the name of the currently active format. This string is not case sensitive.
Setting
Range:
• 'GSM'
• 'AMPS/136'
• 'GPRS'
• 'IS-2000/IS-95/AMPS'
• 'WCDMA'
Query
Range: A string up to 25 characters, or the null string.
• "GSM"
• "AMPS/136"
• "GPRS"
• "IS-2000/IS-95/AMPS"
• "WCDMA"
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SYSTEM:APPLICATION:FORMAT:NAME 'AMPS/136'"
OUTPUT 714;"SYSTEM:APPLICATION:FORMAT:NAME?"
384