Agilent Technologies 8960 Series Reference Manual page 241

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

FETCh:WCPower:POWer[:AVERage]?
Function
Queries the average of the results of several channel power measurements. To set the number of
measurement to average, use the command
Query
• Range: -99 to 99 dBm, 9.91 E+37
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:POWER:AVERAGE?"
!Queries the average power for the channel power results.
FETCh:WCPower:POWer:ALL?
Function
Queries the statistical measurement results for several channel power measurements.
To set the number of measurements to gather statistics for, use the command
"SETup:WCPower:COUNt[:SNUMber]" on page
Results are returned in the following order: minimum, maximum, average, and standard
deviation channel power.
Query
• Range: -99 to 99 dBm, 9.91 E+37
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:POWER:ALL?"
!Queries the statistical measurement results for channel power measurement.
FETCh:WCPower:POWer:MAXimum?
Function
Queries the maximum power of several channel power measurements. To set the number of
measurement to average, use the command
Query
• Range: -99 to 99 dBm, 9.91 E+37
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WCPOWER:POWER:MAXIMUM?"
!Queries the inegrity indicator for the channel power results.
"SETup:WCPower:COUNt[:SNUMber]" on page
304.
"SETup:WCPower:COUNt[:SNUMber]" on page
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_fetch_wcpower.fm
FETCh:WCPower
304.
304..