Agilent Technologies 8960 Series Reference Manual page 235

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

FETCh:WACLeakage:INTegrity?
Function
Queries the measurement integrity indicator for the last adjacent channel leakage measurement
completed.
Zero indicates a normal measurement. See
about non-zero integrity results.
Query
Range: 0 through 23
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WACLEAKAGE:INTegrity?"
!Queries the integrity indicator for the adjacent channel leakage ratio results.
FETCh:WACLeakage:SDEViation[:ALL]?
Function
Queries the standard deviation of several adjacent channel leakage ratio measurements at the
following offsets: -5 MHz, +5 MHz, -10 MHz, +10 MHz.
To set the number of measurement to compare, use the command
"SETup:WACLeakage:COUNt[:SNUMber]"
Query
• Range: 0 to 99 dB, 9.91 E+37
• Resolution: 0.001 dB
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WACLEAKAGE:SDEVIATION:ALL?"
!Queries the standard deviation of power for the adjacent channel leakage ratio
!measurements.
"Integrity Indicator" on page 153
.
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_fetch_wacleakage.fm
FETCh:WACLeakage
for information
235