STATus:OPERation:CALL:COMMon:DATA Condition Register Bit Assignment
The STATus:OPERation:CALL:COMMon:DATA register bits are used to indicate status of processes that
occur during normal data connection processing operations.
Bit
Binary
Number
Weighting
15
32768
14
16384
13
8192
12
4096
11
2048
10
1024
9
512
8
256
7
128
6
64
5
32
4
16
3
8
2
4
1
2
0
1
Condition
Not Used. Defined by SCPI.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Reserved for future use.
Starting Data Connection
Data Connection Status Changing
Data Connection Status Off
Data Connection Status Data
Connected
Data Connection Status
Transferring
Data Connection Status Attached
Data Connection Status Idle
Extension Bit
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is always 0.
This bit is a 1 when:
• Active Cell mode - the data connection status
leaves the idle state
• Test mode - the test set has noted a base station
origination.
This bit is set to a 1 when the data connection
status change detector has been armed.
This bit is set to 1 when the test set is in the call
control status handover/handoff state.
This bit is set to 1 when data connection status data
is in the connected state.
This bit is set to 1 when the data connection status
is Transferring.
This bit is set to 1 when the data connection status
is Attached.
This bit is set to 1 when the data connection status
is Idle.
This bit will always be 0.
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_status_operation.fm