+200 through +299 Error Message Descriptions

Description

A device-specific error indicates that the instrument has detected an error that occurred because some
operations did not properly complete, possibly due to an abnormal hardware or firmware condition. For
example, an attempt by the user to set an out of range value will generate a device specific error.
When one of these errors is generated, the +200 errors bit in the questionable error status register is set.
Table 10. FDD Call Processing Errors
+217 Performance not specified for UL/DL frequency offset
< 30 MHz
+218 FDD Settings Deferred: Cell power and AWGN power
> 10 dB apart.
+221 FDD Test Mode Settings Deferred: Sum of code channel
power levels > 0 dB
+224 FDD Test Mode Settings Deferred: One or more code
channel power levels > 30 dB below total RF power
Error Message
Performance is not specified when the difference in
the uplink and downlink frequencies is less than
30 MHz.
AWGN and Cell Power settings must be within 10 dB
of each other.
The sum of the code channels cannot exceed 0 dB.
Reduce one or more channel levels.
The power of at least one code channel is more than
30 dB below the total RF power, exceeding the test
set's limits.
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_ts_err_msg_pos200.fm