SETup:WTPower:TIMeout:STATe
Function
Sets/queries the thermal power measurement's timeout state.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:WTPOWER:TIMEOUT:STATE ON"
SETup:WTPower:TIMeout:TIME
Function
Sets/queries the thermal power measurement's timeout setting to be used when the timeout state
is set to ON.
Setting
Range: 1 to 999 seconds
Resolution: 1 second
Query
Range: 1 to 999
*RST Setting
10 seconds
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"SETUP:WTPOWER:TIMEOUT:TIME 5S"