Agilent Technologies 8960 Series Reference Manual page 271

Hide thumbs Also See for 8960 Series:
Table of Contents

Advertisement

READ:WBER?
Function
Initiates and fetches a loopback bit error ratio measurement as a sequential operation.
Measurement results are returned in the following order: integrity, bit error ratio, bit error
count, bits tested.
Query
Range
• Integrity: 1 to 23
• Bit Error Ratio: 0 to 99%, 9.91 E+37
• Bit Error Count: 0 to 999999, 9.91 E+37
• Bits Tested: 0 to 999999, 9.91 E+37
Resolution
• Integrity: 1
• Bit Error Ratio: 0.0001 dB
• Bit Error Count: 1
• Bits Tested: 1
Requirements
Test Application Revision: A.02.00 and above
READ:WCPower[:ALL]?
Function
Initiates and fetches a channel power measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average channel power.
If the multi-measurement state is on, the result represents the average of a number of channel
power measurements. Multi-measurement state, by default, is off. To set up
multi-measurements, see
Query
• Range: -99 to 99 dBm
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02.00 and above
"SETup:WCPower:COUNt[:SNUMber]"
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_read.fm
.
READ