READ:AFANalyzer[:ALL]?
Function
Initiates and fetches an audio frequency analyzer measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average audio level, average
SINAD, average audio distortion.
Query
Integrity
• Range: 0 to 23 see
• Resolution: 1
Average audio level
• Range: 0 to 20 volts, 9.91 E+37
• Resolution: the lesser of 4 significant digits or 10 uV.
Average SINAD
• Range: -99.00 to 99.00 dB, 9.91 E+37
• Resolution: 0.01 dB
Average audio distortion
• Range: 0.00 to 99.00%, 9.91 E+37
• Resolution: 0.01%
Requirements
Test Application Revision: A.02.00 and above
READ:WACLeakage[:ALL]?
Function
Initiates and fetches an adjacent channel leakage ratio measurement as a sequential operation.
Measurement results are returned in the following order: integrity, -5 MHz minimum offset
power ratio, +5 MHz maximum offset power ratio, -10 MHz minimum offset power ratio,
+10 MHz maximum offset power ratio.
If the measurement state of any offset is set to OFF, the result for that offset will be 9.91 E+37.
Query
• Range: -100 to 100 dBm
• Resolution: 0.01 dB
Requirements
Test Application Revision: A.02.00 and above