FETCh:WBERror:BITS?
Function
Queries the number of bits tested.
Query
Range: 0 to 999999
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WBERROR:BITS?"
FETCh:WBERror:COUNt?
Function
Queries the bit error count.
Query
Range: 0 to 999999
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WBERROR:COUNT?"
FETCh:WBERror:ICOunt?
Function
Queries the intermediate count of bits tested.
To set up the total number of bits to test, see
Query
Range: 0 to 999999
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WBERROR:ICOUNT?"
FETCh:WBERror:INTegrity?
Function
Queries the measurement integrity indicator for the last loopback bit error measurement
completed.
Zero indicates a normal measurement. See
about non-zero integrity results.
Query
Range: 0 through 23
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"FETCH:WBERROR:INTegrity?"
"SETup:WBERror:COUNt"
"Integrity Indicator" on page 153