SYSTem:UTC

SYSTem
:UTC
"Diagram Conventions" on page 168
SYSTem:UTC[:TIME]
Function
Sets/queries the Universal Coordinated Time (UTC) time of day. (UTC is also known as UCT,
Zulu, and Greenwich Mean Time.)
Local time will track UTC time settings. Also, if new settings are made to local time, UTC time
will track local time.
Setting
Range:
• <hour> 0 to 23
• <minute> 0 to 59
• <second> 0 to 59
NOTE
Query
Range:
• <hour> 0 to 23
• <minute> 0 to 59
• <second> 0 to 59
NOTE
*RST Setting
This feature is not affected by instrument preset
Requirements
Test Application Revision: A.05.05 and above
Programming Example
OUTPUT 714;"SYSTEM:UTC:TIME 7,30,0" !Sets UCT time to 7:30
OUTPUT 714;"SYSTEM:UTC:TIME?" !Returns "+7,+30,+0" if UTC time has been set to 7:30.
[:TIME]
<sp><year>,<month>,<day>
:DATE
? (returns <year>, <month>, <day>)
The resolution of the UTC time setting is minutes. A number within the parameter
range must be entered for <second> but it will not affect the minute setting.
The resolution of the UTC time setting is minutes. The value returned for
<second> will not provide useful information.
<sp><hour>,<minute>,<second>
? (returns <hour>, <minute>, <second>)
S:\Hp8960\E1963A WCDMA\3.2 release\Reference Guide\Chapters\wcdma_hpib_system_utc.fm
SYSTem:UTC