CALL:STATus
CALL:STATus:PICHannel:STATe?
Function
Queries the current state of the downlink page indicator channel.
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:PICHANNEL:STATE?"
CALL:STATus:TOTal:POWer[:AMPLitude]?
Function
Queries the total RF power of the currently selected system type.
The total RF power is equal to the sum of the cell power setting (see
the AWGN setting (see
Query
Range:
• -170dBm/3.84 MHz to +37dBm/3.84 MHz, 9.91 E+37
(-120 dBm/3.84 MHz to -13 dBm/3.84 MHz actual hardware range with amplitude offset = 0)
The actual power range is defined by adding the value of the associated amplitude offset to
the range above. For example, if amplitude offset is -3.5 dB, then the power range will be
-123.5 dBm/3.84 MHz to -16.5 dBm/3.84 MHz.
Resolution: 0.01 dBm/3.84 MHz
*RST Setting
9.91 E+37
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:TOTAL:POWER:AMPLITUDE?"
CALL:STATus:TOTal:POWer:STATe?
Function
Queries the current total RF power measurement's state for the currently selected system type.
This query will return a 1 (on) if the current cell power
on.
Query
Range: 1|0
*RST Setting
0 (off)
Requirements
Test Application Revision: A.02.00 and above
Programming Example
OUTPUT 714;"CALL:STATUS:TOTAL:POWER:STATE?"