Calibration
96
Figure 50
Saving Calibration Data
System Calibration File Structure
Figure 51
shows the files containing system calibration information.
/var/opt/hp93000/soc/calibration/
std__
Figure 51
System Calibration File Structure
std__
CALDrift
cal_log
Auto Calibration Errors
If any errors occurred during Auto Calibration:
1 Record the error messages displayed in the Report window.
2 Print the Auto Calibration log file called:
/var/opt/hp93000/soc/calibration/cal_log
3 For analysis purposes, save a copy of the Auto Calibration configu-
ration file
/var/opt/hp93000/soc/calibration/std__
in a different place.
Note that std__ is a binary file.
Agilent 93000 SOC Series User Training Part 1, October 2004
CalDrift
Contains DC, PMU, AC Linearity and AC Skew calibration
data. Gets overwritten by latest calibration data
D/R Alignment results
Error messages generated during calibration process are
recorded in this file.
5-3 Performing the Auto Calibration