Standard Test Functions
328
7 Select whether you want to make the measurement testing only a
single polarity, or whether you also want to repeat the test with the
opposite polarity. The second test takes place only if the first test
passes.
8 Type your result text in the output entry field. The label $P means
that the result string contains an entry for each pin in the pinlist.
9 Click
exec
started.
Continuity Test Using Programmable Load
This test is used to:
• check the chip bonding and connections to the test fixture,
• check protection circuits in the DUT such as input clamp diodes by
forcing a current into a device pin and measuring the voltage
between that pin and ground,
• check for short-circuits to ground and from pin to pin.
COMMUTATION
COMMUTATION
VOLTAGE
VOLTAGE
DRIVE 0.00 Volts
DRIVE 0.00 Volts
DRIVE 0.00 Volts
DRIVE 0.00 Volts
Figure 208
Continuity Test Using the Programmable Load
Prior to execution of the test, the status of the tester is checked to
determine if it is currently connected or disconnected. If it is
connected, then the DISCONNECT test function is executed. If it is
Agilent 93000 SOC Series User Training Part 1, October 2004
to perform the test or
PROGRAMMABLE
PROGRAMMABLE
LOAD
LOAD
V PASS_MAX
V PASS_MAX
V PASS_MIN
V PASS_MIN
11-4 The Continuity Test Function
break
to abort it once it has
AC RELAY
AC RELAY
AC RELAY
AC RELAY
AC RELAY
AC RELAY