Agilent Technologies 93000 SOC Series Training Manual page 479

User training part 1
Hide thumbs Also See for 93000 SOC Series:
Table of Contents

Advertisement

17-2 Pin Margin Setup
Default Settings
Agilent 93000 SOC Series User Training Part 1, October 2004
Specifying the Settings
The following sections describe how to set up the various settings for
each type of margin measurement. You specify the individual settings
in the pulldown menu Settings.
Figure 289
Pin Margin Tool Settings Menu
Unless you specifically specify settings for a pin margin test, the
defaults are used. These default pin margin settings are as follows:
• All used edges, pins, timing & levels sets, and level resources (vih,
vil, voh, vol, vch, vcl, vt, ioh, iol) will be used.
• no period will be used
• the current testsuite (dataset, label, testfunction and primary sets)
and result focus site will be used
• there will be no DPS variations
• parallel evaluation will be used — that is, all pins with the same
setup settings are measured together
• the edge delay range is relative to the actual settings (±5% of period)
• the period range is ±1% of the actual period
• the level range of ±0.2V is relative to actual settings
• load value of 0 to 30mA (absolute values)
How to change the default settings is described in the subsequent
modules.