Timing Setup
Creating the Waveforms
240
– drag the edge to the desired position in the tester cycle area.
or
– enter the edge position in the edge delay overlay of the Timing
Entry Area.
10 Repeat this for all edges needed.
The third step when setting up the timing is to assign shapes to the
edges to create waveforms:
11 Select the edges, to which you wish to assign a shape (one, two, or
three edges depending on the shape).
Use click and shift-click operations to select the edges.
12 Select
Select -
You see the Shapes window. The Shapes Window allows you to
change between drive and receive shapes with the
In
Figure 156
Figure 156
Drive Shapes Window
In
Figure 157
Agilent 93000 SOC Series User Training Part 1, October 2004
shapes....
you see the drive shapes.
you see the receive shapes.
9-9 Fixed Timing