Pin Margin Algorithm
594
This procedure is represented schematically in
Step 1
+ve limit
actual value
-ve limit
+ve and -ve limits
set by user
Figure 330
How the Margins are Determined
'Only Test to First Transition' selected
Another (quicker) way of performing your pin margin test is to select
the option Only Test to First Transition in the window Mode Settings
(Settings pulldown menu).
The principle is the same as that outlined in the previous section,
except that the starting point of the search is the actual value now. A
measurement is performed starting from the actual value in the direc-
tion of the lower limit until a fail is encountered. A measurement is
then performed starting from the actual value in the direction of the
upper limit until a fail is encountered. A binary test is then performed
at each of these transition points. This way only the points in a pass
region are measured.
The number of steps can be chosen between 2 and 50. The default
number of steps is 50.
This technique is shown in
Agilent 93000 SOC Series User Training Part 1, October 2004
Step 2,3
actual
fail
minimum
cycle (FFC)
Step 3 i
Step 3 ii
binary search
for lower limit
Figure 331
(p. 595).
Figure
330.
Step 4
maximum
cycle (FFC)
actual
fail
Step 4 ii
Step 4 i
binary search
for upper limit