MARKer:LEVel <numeric_value>
This command sets the marker measurement level.
Prior to sending this command, you need to set the marker measurement mode
NOTE
to DELay or SPACing (refer to
This command sets the marker level for both marker A and marker B.
Syntax
Parameter
Item
numeric_value
Keysight 8990B Programming Guide
MARK
:LEV
Description/Defaul t
A numeric value in % for the marker measurement level:
– Default value: 50%
– Minimum value: Reference level 1
(refer to
"MEASure:THReshold:REFlevel{1|2}
CHAN{1|2|3|4},<numeric_value>"
channel-dependent)
– Maximum value: Reference level 2
(refer to
"MEASure:THReshold:REFlevel{1|2}
CHAN{1|2|3|4},<numeric_value>"
channel-dependent)
"MARKer:MODE <character_data>"
Space
numeric_value
?
on page 265,
on page 265,
MARKer Subsystem
on page 321).
Range of values
Reference level 1 to reference
level 2
11
313