LIMIT:RUNop <character_data>, [numeric_value]
This command determines the duration of the limit test. The limit test can be set
to run forever, or to stop after a certain duration or after a set number of
waveform captures.
Syntax
LIMIT
Parameter
Item
character_data
numeric_value
Example
LIMIT:RUN FOR
Keysight 8990B Programming Guide
:RUN
Description/Defaul t
Sets the test operation type to:
DURation
FORever
WAVeforms
The range of numeric values is only valid for the DURation and
WAVeforms test operation types:
For the DURation test operation type:
0.1 to 60 minutes
For the WAVeforms test operation type:
0 to 100k
Space
character_data
This command sets the test operation type to run
forever.
LIMIT Subsystem
numeric_value
,
Range of values
DURation
FORever
WAVeforms
0.1 - 60 minutes
0 - 100k