Acquire:rf:interpolate:threshold <Character_Data - Keysight 8990B Programming Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

4
ACQuire Subsystem
ACQuire:RF:INTerpolate:THReshold <character_data>
This command is used to specify the timebase threshold. The interpolation
algorithm will start to interpolate trace data acquired by the 8990B when the
timebase falls below the specified threshold.
This command can only be used when ACQuire:RF:INTerpolate is set to AUTO
NOTE
(refer to
Syntax
Parameter
Item
character_data
Example
ACQ:RF:INT:THR THR2U
Reset condition
On reset, the timebase threshold is set to THR500N.
136
""
on page 141).
ACQ
:RF
Description/Defaul t
Sets the timebase threshold to:
– THR500N: Sets the threshold to 500 nanoseconds per division.
– THR1U: Sets the threshold to 1 microsecond per division.
– THR2U: Sets the threshold to 2 microseconds per division.
– THR5U: Sets the resolution to 5 microseconds per division.
– THR10U: Sets the resolution to 10 microseconds per division.
– THR20U: Sets the resolution to 20 microseconds per division.
:INT
:THR
Sets the timebase threshold to 2 microseconds
per division.
Keysight 8990B Programming Guide
Space
character_data
Range of values
THR500N
THR1U
THR2U
THR5U
THR10U
THR20U