3
Root Commands
ETS <character_data>
This command sets the state of the Equivalent-Time Sampling (ETS) event.
Syntax
Parameter
Item
character_data
Example
ETS OFF
Remarks
– Turning on ETS enables the 8990B to capture high-frequency signals with a
– Video bandwidth is turned off when ETS is enabled.
– The 10 MHz reference signal input is turned off when ETS is enabled. If ETS is
– During AUX trigger mode, pressing the single button will ensure that ETS
68
Space
ETS
Description/Defaul t
Sets the ETS state as:
– AUTO: Sets ETS to automatically turn on when the timescale
value is less than the value set at ETSThreshold.
– OFF: Turns off ETS.
This command turns off the ETS event.
higher effective sampling rate.
enabled while the 10 MHz input is turned on, error 939, "10 MHz setting is only
applicable when ETS is off. 10 MHz setting is turned off" will occur.
acquisition is completed before returning the trace. As compared to other
trigger sources, the user needs to send multiple single commands or press the
single button to get the completed ETS trace.
character_data
?
Range of values
AUTO
OFF