TRIGger:FAST <boolean>
This command sets the state of the fast trigger.
Syntax
TRIG
Example
TRIG:FAST ON
Remarks
– This command is only applicable for the channels 1 and 4 only. If this
– Auto-calibration will be turned off when the fast trigger is enabled.
Reset condition
On reset, the fast trigger is disabled.
Keysight 8990B Programming Guide
:FAST
This command enables the fast trigger.
command is sent for a trigger source other than any of these sources, error
–221,"Settings conflict;Fast trigger is only applicable to trigger source for
channel 1 and 4. Fast trigger has been disabled" will occur.
Space
0|OFF
1|ON
?
TRIGger Subsystem