FUNCtion{1|2|3|4}:COMMonmode
<character_data>,<character_data>[,<numeric_value>]
This command sets the source for the common mode operation. The common
mode operation adds two sources and divides the resultant by two. It is done
point by point.
– If the source(s) input is in the log scale, it will be converted to the linear scale
NOTE
– If the trace length between the channels is not equal, the waveform math
– If the number of trace points is not equal, the waveform math operation will
Syntax
1
FUNC
2
3
4
338
to perform the common mode math operation. The resultant trace is then
converted back to dBm.
operation will find the shortest trace between the sources and shrink other
traces to an equal length.
add trace points by using the PCHIP method.