Acquire:rf:interpolate:resolution <Character_Data - Keysight 8990B Programming Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

ACQuire:RF:INTerpolate:RESolution <character_data>
This command sets the resolution during interpolation in number of points per
nanosecond.
This command can only be used when ACQuire:RF:INTerpolate is set to AUTO
NOTE
(refer to
Syntax
Parameter
Item
character_data
Example
ACQ:RF:INT:RES POINT2N
Reset condition
On reset, the interpolation algorithm resolution is set to POINT1N.
Keysight 8990B Programming Guide
""
on page 141).
ACQ
:RF
Description/Defaul t
Sets the interpolation algorithm resolution to:
– POINT1N: Sets the resolution to 1 point per nanosecond.
– POINT2N: Sets the resolution to 2 points per nanosecond.
– POINT5N: Sets the resolution to 5 points per nanosecond.
:INT
:RES
Sets the interpolation algorithm resolution to 2
points per nanosecond.
ACQuire Subsystem
Space
character_data
Range of values
POINT1N
POINT2N
POINT5N
4