CALibration:ZEROCAL <character_data>
This command sets the 8990B to perform zeroing and calibration on channel 1 or 4.
Syntax
Parameter
Item
character_data
Example
CAL:ZERO CHAN1
Remarks
– This command is only applicable for channels 1 and 4. If a channel other than
If channel 1 or 4 is not connected, error –241,"Hardware missing; Sensor not
found in channel 1 or channel 4." will occur.
Keysight 8990B Programming Guide
CAL
:ZEROCAL
Description/Defaul t
Sets the channel for zeroing and calibration as:
– CHAN1: Sets channel 1 to zero and runs calibration.
– CHAN4: Sets channel 4 to zero and runs calibration.
channel 1 or 4 is specified, error 700,"Applicable to channel 1 and 4 only" will
occur.
Space
character_data
This command sets the 8990B to perform zeroing and
calibration on channel 1.
CALibration Subsystem
Range of values
CHAN1
CHAN4
5
155