4
ACQuire Subsystem
ACQuire:CCDF:MARKer:BSOUrce <character_data>
This command sets the CCDF marker B source.
Syntax
ACQ
Parameter
Item
character_data
Example
ACQ:CCDF:MARK:BSOU TRAC4
Remarks
– This command is only applicable for channels 1 and 4. If you send this
– This command is only applicable in the CCDF acquisition mode. If the
102
:MARK
:CCDF
Description/Defaul t
Sets the source of marker B as:
– TRACe1: Places marker A on channel 1 CCDF trace.
– TRACe4: Places marker A on channel 4 CCDF trace.
– GAUSsian: Places marker A on Gaussian trace.
– REFerence: Places marker A on reference trace.
– NONE: Disables marker B.
command for a channel other than 1 or 4, error 700,"Applicable to channel 1
and 4 only" will occur.
acquisition mode is not set to CCDF, error –221,"Settings conflict;Requires
Space
:BSOU
This command sets marker B on
channel 4 CCDF trace.
Keysight 8990B Programming Guide
character_data
?
Range of values
TRACe1
TRACe4
GAUSsian
REFerence
NONE