CALibration:CALibrate:AUTO:INTerval <numeric_value>
This command sets the time interval to run auto calibration on channels 1 and 4.
Syntax
Parameter
Item
numeric_value
Example
CAL:CAL:AUTO:INT 1
CAL:CAL:AUTO:INT 9.5
Keysight 8990B Programming Guide
CAL
:AUTO
:CAL
Description/Defaul t
A numeric value in minute for the time interval:
– Default value: 10 min
– Minimum value: 1 min
– Maximum value: 60 min
CALibration Subsystem
:INT
Space
This command sets the auto calibration to run
on a 1 min interval.
This command rounds up 9.5 to the nearest
integer, and auto calibration will run on a 10 min
interval.
numeric_value
?
Range of values
1 min to 60 min
(decimals will be rounded to
the closest integer)