Trigger:impedance <Character_Data - Keysight 8990B Programming Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

TRIGger:IMPedance <character_data>
This command sets the impedance for the auxiliary trigger source input.
Syntax
Parameter
Item
character_data
Example
TRIG:IMP IMP50
Remark
This command is only applicable for the auxiliary trigger source input. If this
command is sent for a trigger source other than auxiliary, error 704,"Applicable to
trigger source for auxiliary only" will occur.
Reset condition
On reset, the impedance is set to 50 Ω.
Keysight 8990B Programming Guide
TRIG
:IMP
Description/Defaul t
Sets the impedance type as:
– IMP50: Sets the impedance to 50 Ω. This is the default setting.
– IMP1M: Sets the impedance to 1 MΩ.
Space
character_data
?
This command sets the auxiliary trigger source input
Ω
impedance to 50
.
TRIGger Subsystem
Range of values
IMP50
IMP1M
8