Example 2: Making Pulse Characterization Measurement - Keysight 8990B Programming Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

Example 2: Making Pulse Characterization Measurement

The following command sequences provide the example to perform the pulse
characterization measurement.
"–>" indicates the commands that you send to the 8990B.
NOTE
This measurement example requires the following test setup:
E4438C signal generator
Figure 23-1
E4438C signal generator configuration
-> SYST:PRES
Wait 1 sec for operation to complete.
-> *CLS
-> FREQ 1GHZ
-> POW 0DBM
-> PULM:INT:PER 1000uS
-> PULM:INT:PWID 200uS
-> PULM:STAT ON
Keysight 8990B Programming Guide
N1923/24A sensor
Test setup for the pulse characterization measurement
//Presets the E4438C to its default settings.
//Clears the error queue.
//Sets the frequency to 1 GHz.
//Sets the output power level to 0 dBm.
//Sets the pulse period to 1000 μs.
//Sets the pulse width to 200 μs.
//Turns on the pulse signal.
Programming Examples
8990B peak power analyzer
23
549

Advertisement

Table of Contents
loading

Table of Contents