Calibration:zero <Character_Data - Keysight 8990B Programming Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

5
CALibration Subsystem
CALibration:ZERO <character_data>
This command sets the 8990B to perform zeroing on channel 1 or 4.
Syntax
Parameter
Item
character_data
Example
CAL:ZERO CHAN1
Remarks
– This command is only applicable for channels 1 and 4. If a channel other than
– If channel 1 or 4 is not connected, error –241,"Hardware missing; Sensor not
154
CAL
:ZERO
Description/Defaul t
Sets the channel for zeroing as:
– CHAN1: Sets channel 1 to zero.
– CHAN4: Sets channel 4 to zero.
channel 1 or 4 is specified, error 700,"Applicable to channel 1 and 4 only" will
occur.
found in channel 1 or channel 4." will occur.
Space
character_data
This command sets the 8990B to perform zeroing on
channel 1.
Range of values
CHAN1
CHAN4
Keysight 8990B Programming Guide

Advertisement

Table of Contents
loading

Table of Contents