5
CALibration Subsystem
CALibration:OUTPut <character_data>
This command sets the AC coupling frequency or DC level of the calibrator
waveform output through the rear panel Auxiliary Trigger Out connector.
Syntax
Parameter
Item
character_data
Example
CAL:OUTP DC,2.0
Remark
The AC setting sets the Aux Trig Out to be the probe compensation square wave
(approximately 750 Hz).
152
CAL
:OUTP
Description/Defaul t
Sets the calibration output as:
– AC: Sets the AC coupling frequency.
– DC,<dc_value>: Sets the DC level, with its value in volts
within the range of –2.4 V to +2.4 V.
character_data
Space
?
This command puts a DC voltage of 2.0 V on the Aux
Trig Out connector..
Keysight 8990B Programming Guide
Range of values
AC
DC,<dc_value>