ACQuire Subsystem

ACQuire:CCDF:PEAK?

This query returns the CCDF peak power value (in dBm) for the selected CCDF
source.
Syntax
Example
ACQ:CCDF:PEAK?
Remarks
– This query is only applicable for channels 1 and 4. If you send this query for a
– This query is only applicable in the CCDF acquisition mode. If the acquisition
– This query is only applicable when the CCDF source is set. If the source is not
110
ACQ
:CCDF
channel other than 1 or 4, error 700,"Applicable to channel 1 and 4 only" will
occur.
mode is not set to CCDF, error –221,"Settings conflict;Requires CCDF mode to
be enabled" will occur. Refer to
page 132 to set the acquisition mode.
set, error –221,"Settings conflict;CCDF source not set" will occur. Refer to
"ACQuire:CCDF:SOURce <character_data>"
source.
:PEAK
?
Queries the CCDF peak power value for selected CCDF
source.
"ACQuire:MODE <character_data>"
on page 117 to set the CCDF
Keysight 8990B Programming Guide