19
MEMory Subsystem
MEMory:TABle:EDIT <numeric_value>,<numeric_value>
This command modifies the offset value of an existing frequency and offset pair in
the FDO table selected in
Syntax
MEM
Parameter
Item
numeric_value
Example
MEM:TAB:EDIT 1000,30
486
:EDIT
:TAB
Description/Defaul t
Numeric values for the frequency and offset pair:
– Minimum values:
– Frequency: 1 kHz
– Offset: 1%
– Maximum values:
– Frequency: 1 THz
– Offset: 150%
The first <numeric_value> represents the frequency while the
second <numeric_value> represents the offset. The frequency
value specified will be truncated to a multiple of 1000.
"MEMory:TABle:SELect <string>"
Space
numeric_value
This command changes the offset to 30% for the
1 kHz frequency in the selected FDO table.
Keysight 8990B Programming Guide
on page 494.
numeric_value
,
Range of values
1 kHz frequency, 1% offset to 1
THz frequency, 150% offset