TRIGger:HYSTeresis:MODE <character_data>
This command sets the trigger hysteresis mode for channels 2 and 3.
Syntax
Parameter
Item
character_data
Example
TRIG:HYST:MODE NORM
Remark
This command is only applicable for channels 2 and 3. If you send this command
for channels other than 2 and 3, error 706,"Applicable to trigger source for
channel 2 and 3 only" will occur.
Reset condition
On reset, the trigger hysteresis mode is set to normal sensitivity.
Keysight 8990B Programming Guide
TRIG
:HYST
Description/Defaul t
Sets the hysteresis mode as:
– NORMal: Sets the mode to normal sensitivity. This is the default
setting.
– NREJect: Sets the mode to noise reject.
Space
:MODE
This command sets the trigger hysteresis mode to
normal sensitivity.
TRIGger Subsystem
character_data
?
Range of values
NORMal
NREJect