Figure 4-8. Standard Event Status Register - Agilent Technologies N9360A Programming Manual

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

4
Programming Command Guidelines
Table 4-5 Status Byte Register
Bit
Description
0
If all pending operations is completed following execution of the *OPC command, this bit is set to 1.
1
This bit is always set to 0.
2
If a query error has occurred, this bit is set to 1.
3
If a device dependent error has occurred, this bit is set to 1. Device dependent errors have error numbers from
-399 to -300 and 1 to 32767.
4
If an execution error has occurred, this bit is set to 1.
Execution errors have error numbers from -299 to -200.
5
If a command error has occurred, this bit is set to 1.
Command errors have error numbers from -199 to -100.
6
This bit is always set to 0.
7
If the instrument has been turned off and then on, this bit is set to 1.
4-20
Standard Event Status Register
The Standard Event Status Register of the Tester is used to
determine the specific event that set bit 5 in the status byte
register as follows:
Figure 4-8 Standard Event Status Register
Bit Number
7
6
The Standard Event Status Register contains the following
information:
To query the Standard Event Status Register, send the
command *ESR?. The response is the decimal sum of the bits
which are set to
. For example, if the bit 7 and bit 3 are set to
1
the decimal sum of these 2 bits is 136 (128 plus 8) see
Figure
4-5. So the decimal value 136 is returned.
N9360A Multi UE Tester CDMA2000 Programming Manual
5
4
3
2
*
ESR?
1
0
,
1