Agilent Technologies N9360A Programming Manual page 177

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

N9360A Multi UE Tester CDMA2000 Programming Manual
• <real 3> (up to 6 letters) is the measured data ranging
from -99.99 to +99.99 in dBm. If measurement result is
out of range or MIN TX Power is not measured, the value
+99.99 is returned.
Frequency Error Measurement
TESTs:MANual:RESults:FERRor?
This query command returns the Frequency Error Measurement
result values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Frequency Error is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 8 letters) is the measured data ranging
from -10000.0 to +10000.0 in Hz. If measurement result is
out of range or Frequency Error is not measured, the value
+99999.9 is returned.
Rho Measurement
TESTs:MANual:RESults:RHO?
This query command is valid when CDMA mode is set to MC1X.
This returns the Rho Measurement result values, <int 1>, <int
2>, and <real 3> when F1R1 or F2R2 is set by the Radio
Configuration command.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Rho is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 5 letters) is the measured data ranging
from 0.800 to 1.000. If measurement result is out of range
or Rho is not measured, the value 9.999 is returned.
Programming Command Reference
5
5-127