Programming Command Reference
5-146
Time Offset Measurement
TXANalyzer:RESults:TOFFset?
This query command returns the Time Offset Measurement result
values, <int 1>, <int 2>, and <real 3>. Each value is
separated by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement range is out of
range or Time Offset is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 6 letters) is the measured data ranging
from -10.00 to +10.00 in
of range or Time Offset is not measured, the value 99.99 is
returned.
Origin Offset Measurement
TXANalyzer:RESults:OriginOFFset?
This query command returns the Origin Offset Measurement result
values, <int 1>, <int 2>, and <real 3>. Each value is
separated by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement range is out of
range or Origin Offset is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 4 letters) is the measured data ranging
from 0.0 to 99.9 in dBc. If measurement result is out of
range or Origin Offset is not measured, the value 99.99 is
returned.
Code Domain Power (TX Power) Measurement
TXANalyzer:RESults:CDPWr:TXPower?
This query command returns the TX Power measurement result
values, <int 1>, <int 2>, and <real 3> on the [Code
Domain Power] measurement screen. Each value is separated
by a comma.
N9360A Multi UE Tester CDMA2000 Programming Manual
μ
sec. If measurement result is out